7 - 25
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
7.10.3.2 Graphically Programming Integration Events
1
Right-click on the chromatogram and select the desired integration event under
Graphical Programming
.
2
A blue window will display (see
). Using the pointer, select a position
for the
Start Time
and the
Stop Time
on the chromatogram for the specific
event.
Figure 7-30 Blue window for graphical programming
3
Once
Start Time
and
Stop Time
are selected, a window will display with the
times (see
). Verify the
Start Time
,
Stop Time
, and
Value
. Click
Analyze Now
to close the window.
Figure 7-31 Example window for graphical programming
All events added through
Graphical Programming
will be added to the
method
Integration Events
table by default. There is an option to
Add event
to this data file only
. Refer to
4
To save the method, click
File >> Method >> Save
.