8 - 24
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
8.3.2 Running a Low Pressure, Clean Gas Sample
To obtain a low pressure, clean gas sample, use the standard 10 micron sample
inlet filter assembly for filtration (PN G2801-60900, included in the ship kit).
1
Turn 3000 Micro GC on.
2
Make sure the carrier gas supply is sufficient to run all of the samples.
3
If using a sample vessel, prepare the sample.
NOTE:
If not installed, install the sample inlet filter assembly. Refer to
4
Connect the sample container or gas source to the 3000 Micro GC sample
inlet.
NOTE:
For sample source pressures above 172 kPa (25 PSI), install a pressure
regulator as shown in
Figure 8-29 Example sample connection for a clean, low pressure sample
5
Open the sample vessel valve, and adjust the pressure to <172 kPa (<25 PSI).
6
Use EZ IQ software to load the desired method and perform the analysis.
7
When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.
8
Repeat steps 3 to 7 to run the next sample.
Set input pressure
0 to 172 kPa
(0 to 25 PSI)
Recommended
35 to 70 kPa
(5 to 10 PSI)
Safety stand
for cylinder
Dual-end ferrule
External filter assembly
G2801-60900
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...