6 - 11
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
8
Select the new carrier gas type from the drop-down list. See
Figure 6-13 Selecting a new carrier gas type
9
Click
Submit
. This turns the column heaters off (to cool the columns and avoid
thermal shock) and displays the
Check Status
view. See
Figure 6-14 Check status view
10
Click
Check Status
to display a window which shows the current column
temperature(s). When the column temperatures are below 60
º
C,
Close
the
Check Status
window. See
NOTE:
Changes may be canceled up to this point by closing the browser or by
using the browser back feature.
Figure 6-15 Example check status window
Periodically refresh display
Actual column tempertures