3000 Micro GC Operating Manual
8.2.4 Decommission a Standard GC Module
A Standard GC module must be decommissioned within the embedded web page
before it can be removed. 3000 Micro GC internally communicates to each
installed GC module using a unique module address (see
).
For a 1-2 channel or portable 3000 Micro GC, the module addresses are
1 and 2.
For a 3-4 channel 3000 Micro GC, the module addresses are 1, 2, 3 and 4.
Decommissioning a standard GC module will disable the GC module's address.
Figure 8-3 Default GC module serial communication addresses
NOTE:
For Performance Enhanced modules, skip to
NOTE:
Only one Standard GC Module can be decommissioned at a time.
The procedure below describes the procedure for replacing the Standard GC
module 1 (Channel A) in a 2-channel 3000 Micro GC. The process is similar for
other configurations.
1
Open the web browser and type in the 3000 Micro GC IP Address.
The embedded webpage will display.
NOTE:
At the factory, the instrument IP address was set to
10.1.1.101
.
2
Click
the
Utilities
tab.
3
Click
Change Module Configuration
. See
Address
1
1
2
2
3
4
Front
Front
2-channel 3000
Micro GC (top down view)
4-channel 3000
Micro GC (top down view)
Module
address
Corresponding
channel
1
2
3
4
A
B
C
D
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...