3000 Micro GC Operating Manual
Figure 8-1 Example bakeout method
3
Click
File >> Method >> Save
to save the bakeout method. When prompted,
type a unique name for the bakeout method.
4
Verify that the carrier gas is on and flowing. This will protect the column and
TCD.
5
Click
Control >> Download Method
to download the bakeout method.
6
Allow the method to run for the duration listed in
.
7
After the bakeout is complete, load an analytical method and run a set of
calibration samples. Refer to
.
8
Check the report. Adjust the calibration settings, retention times, and method
settings as needed.