7 - 36
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
7.12 Calibration
A calibration must be performed to quantify a peak (i.e., determine the amount of
a sample based on known standard gases). In this section, the procedure for
setting up single level and multiple level calibration is described.
7.12.1 Obtain Preliminary Data for Calibration
To enter calibration peak data, run a sample of a calibration gas (i.e., a sample with
known component concentrations). The generated data file can be used to define
the calibration peaks.
1
Click
Control >> Single Run
to run the sample.
2
Type the desired
Sample ID
and click
Start
.
3
After the run is complete, ensure all peaks are integrated and annotated.
7.12.2 Enter Concentration Amounts in the Peaks/Groups Table
The known concentration of components in the calibration gas must be recorded in
the
Peaks/Groups
table.
NOTE:
If calibration concentrations were entered when defining single peaks
(see
), this step can be skipped.
1
Click
Method >> Peaks/Groups
.
2
Use the scroll bar to scroll until the
Level 1
column is visible. See
.
Figure 7-46 Level information in the peak table
3
Type the concentration of the component.
4
Repeat this for all peaks and groups in each of the channels.
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...