9 - 13
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
9.2.1.2 Retention Time Symptoms
9.2.1.2.1 Retention Time Drift
Retention time drift is a steady increase or decrease of retention times in a series
of successive runs. Erratic times (both directions) are discussed later as retention
time wander.
Retention times suddenly increase significantly
Decreased carrier flow or reduced column temperature; check the pressure
and temperature set points
The carrier gas tank may be nearly empty
Retention times suddenly decrease significantly
Increased carrier flow or increased column temperature; check the pressure
and temperature set points
9.2.1.2.2 Retention Time Wander (Reproducibility)
Reproducibility is erratic for successive runs of similar composition
Temperature or pressure variations within the runs. May indicate possible
detector or control failure
Radical differences in sample component concentrations alter retention time
Reproducibility is acceptable later in the run but not for early eluting peaks
When the earliest peaks elute very rapidly, they may not achieve chromatographic
equilibrium with the stationary phase. They act as solvent peaks and are carried
straight through the column.
Lower the column temperature.
NOTE:
A 30°C drop approximately doubles the retention time.
Retention time changes with large sample concentrations
When there is more sample than the column can handle, peaks will be deformed
and will shift to earlier retention times. Dilute the sample or shorten
Inject Time
.
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...