3000 Micro GC Operating Manual
2
Turn 3000 Micro GC on.
3
Wait three minutes until 3000 Micro GC has fully powered up.
4
Use EZ IQ software to load the desired method.
5
Power on the Heated Vaporizer.
6
Wait approximately five minutes for the Heated Vaporizer Assembly to stabilize
at operating temperature.
7
Make sure the sample vessel stopcock is closed and the relief valve is turned
fully to
Sample
.
8
Connect the sample vessel to the filter disconnect assembly on the Heated
Vaporizer. See
Figure 8-35 Sample vessel installed on a heated vaporizer (2-channel instrument shown)
9
Open the stopcock on the sample vessel.
10
Slowly turn the relief valve toward
Vent
until a small but steady flow is vented
and allow the sample to purge the line for approximately thirty seconds.
11
Turn the relief valve to
Sample
.
12
Use EZ IQ software to load the desired method and perform the analysis.
13
When the analysis is complete, close the stopcock on the sample vessel.
Sample vessel
Sample vessel stop-cock
Filter disconnect assembly
Quick disconnect fitting
Low pressure vent tubing
High pressure vent tubing
Relief valve
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...