P1
C
3000 Micro GC Operating Manual
7
A window will prompt to restart 3000 Micro GC. Click
Restart
and wait at least
three minutes before shutting down the 3000 Micro GC. See
CAUTION
Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.
Figure 6-21 IP configuration - restart
8
Turn off 3000 Micro GC.
9
If needed, reconfigure the computer’s IP address for local LAN use.
10
If needed, disconnect the crossover cable, and connect the 3000 Micro GC and
the computer to the local LAN using standard LAN cables (PN G1530-61495).
11
If needed, reboot the computer.
12
Turn on 3000 Micro GC.
13
Reconnect to 3000 Micro GC using the new IP address. Ping the new IP
Address. For example, if the new IP address is 10.1.1.102, click
Start >> Run
and type
ping 10.1.1.102.
Click
OK
Figure 6-22 Ping new IP address