7.15.2 Running a Sample with a Valco Valve
1
In
EZ IQ
, click
Control >> Single Run
to display the
Single Run Acquisition
window. See
.
Figure 7-69 Single run acquisition window with valve control
2
Type a
Vial
number. This is the valve port position for the current single run.
3
Type an
Injection volume
. This is the
line purge time
for the current single run.
NOTE: Injection volume
is labeled as
uL
. However, the value corresponds to
time in minutes.
Use the above settings as an example:
Prior to the start of a run, the valve will switch to the position corresponding
to the
Vial
number,
2
in this case. After the time defined in the
Injection
volume
,
1
minute in this case, the run will start.
After the run has started, the valve will switch back to the
Valve purge
position
after the
Purge delay
time specified on the
Sampling
tab in
Method >> Instrument Setup
. See