3000 Micro GC Operating Manual
ISTD ID#
If an internal standard calibration will be performed, enter
the ID # for the internal standard peak for this compound.
The ID # refers to the peak
ISTD. ID #
from the
Peaks/Groups
table. If the
ISTD. ID #
is unknown, the
information can be added in the peak table later.
Ref ID#
Enter retention time reference peak ID # to be used for this
peak. The Ref ID # refers to the peak
Ref. ID #
from the
Peaks/Groups
table. If the
Ref. ID #
is unknown, the
information can be added later in the peak table. Reference
peaks are used to locate calibrated peaks when
chromatographic conditions change such that retention
times shift.
Retention time window
Select either
Relative
or
Absolute Retention time
window
.
The
Relative
option is based on a percent of the expected
retention time of the peak.
The
Absolute
option is based on a specified value.
The retention time window compensates for slight
deviations from the expected retention time.
Table 7-3 Parameters
Parameters
Description
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...