3000 Micro GC Operating Manual
9.2.1.3 Peak Symptoms
9.2.1.3.1 No Peaks
Possibilities include operator error, insufficient
Sample Pump
time, low
Detector
Sensitivity
setting (peaks are present but not visible) or insufficient
Inject Time
.
Check the 3000 Micro GC analytical and reference vent flow. No flow could
indicate a broken column. The presence of flow could indicate an injector or TCD
malfunction. Contact
.
9.2.1.3.2 Inverted Peaks
Inverted peaks are usually caused by poor quality carrier gas. The carrier gas must
be 99.995% to 99.9995% pure. Inverted peaks may indicate that the carrier gas is
contaminated, or that the incorrect carrier gas is being used, or it is configured
improperly.
Carrier gas configuration can be checked on the 3000 Micro GC embedded web
page. Refer to
. If the carrier gas is configured incorrectly,
correct it by following the procedure outlined in
for carrier gas requirements.
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...