6.3 ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 11
LifeCare PCA
636
Motor winding excessive current
Motor winding current
above 200 mA
Replace mechanism
assembly
If problem persists,
637
Motor winding open circuit
Motor winding current
below 75 mA
638
Motor Voltage Error
Motor voltage out of range
639
Contrast Voltage Error
Contrast voltage out of range
Verify LCD assembly
connections to MCU PWA
Replace LCD assembly
If the problem persists,
640
Internal RAM Error
Failed read/write operations
on internal RAM
641
Occlusion Switch Error
Defective occlusion sensor
Adjust occlusion
alarm switch
Replace mechanism
assembly
If problem persists,
642
Empty Syringe Switch Error
Defective empty
syringe sensor
Replace mechanism
assembly
If problem persists,
643
Injector Switch Error
Defective injector sensor
Adjust injector switch
Replace mechanism
assembly
If problem persists,
644
Watchdog Timer Error
Watchdog failure
Verify jumper on pins
5 - 6 of JP1 on MCU PWA
If problem persists,
645
Barcode Reader Error
Three consecutive attempts
to read the barcode have
failed
Replace barcode reader
If problem persists,
646
UART Loopback Test Error
Serial communication
chip failure
Table 6-2. Error Codes Requiring Technical Service
Alarm
Code
Malfunction Description
Possible Cause
Corrective Action