SECTION 6 TROUBLESHOOTING
LifeCare PCA
6 - 14
Technical Service Manual
Delivery Accuracy Test
LifeCare PCA set not properly
primed
Re-prime set and check clamps
Defective mechanism
assembly
Replace mechanism assembly
Occlusion Test
Set not properly primed
or installed
Prime set
Defective mechanism
Replace mechanism assembly
Occlusion alarm switch
requires adjustment
Electrical Safety Test
Defective AC power cord
Replace AC power cord
Table 6-3. Troubleshooting with the PVT
Test Failure
Probable Cause
Corrective Action