6.3 ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 7
LifeCare PCA
313
Battery Current Error
Battery charge current
exceeded limit
Verify battery connections
Replace battery
480
Configuration Data
Failed CRC of pump
configuration data
For each malfunction
press
[CLEAR]
,
then
[ON/OFF]
to power
off the infuser,
then
[ON/OFF]
to power
on the infuser;
continue until all
malfunctions are cleared
Verify lithium coin cell
battery connections
Verify infuser is connected
to AC power
Replace lithium coin cell
battery
Replace battery
If malfunction cannot be
cleared,
481
Infusion Data
Failed CRC of infusion data
482
Program Data
Failed CRC of program data
483
Flash Checksum Error
Failed checksum at startup
or during operation
484
Stored Protocol Data
Failed CRC of stored
protocols
485
Motor State
Invalid rate or restart;
bad states or events
in motor software
486
History Corruption
History data corrupted
487
Motor Data Error
Failed CRC of motor data
488
Drug Table Data Error
Failed CRC of drug table data
Table 6-2. Error Codes Requiring Technical Service
Alarm
Code
Malfunction Description
Possible Cause
Corrective Action