SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 40
Technical Service Manual
10. Verify the
ENTER DRUG CONCENTRATION
message displays. Using the numeric
keypad, enter 1, then press
[ENTER]
.
11. Verify the
CONFIRM 1 mg/mL CONCENTRATION TO PHYSICIAN RX
message displays,
then press
[CONFIRM]
.
12. Verify the
SET LOADING DOSE?
message displays, then press
[YES]
.
13. Verify the
ENTER LOADING DOSE
message displays. Using the numeric keypad,
enter 10 mg, then press
[ENTER]
.
14. Verify the
SELECT DELIVERY MODE
message displays. Press
[CONTINUOUS]
.
15. Verify the
ENTER CONT RATE
message displays. Using the numeric keypad, enter
20 mg/Hr, then press
[ENTER]
.
16. Verify the
SET A DOSE LIMIT
message displays. Press
[NO]
.
17. Verify the
CONFIRM NO DOSE LIMIT
message displays. Press
[CONFIRM]
.
18. Verify the
LOADING DOSE
message displays. Press
[CONFIRM]
.
19. Verify the
CONT RATE
message displays. Press
[CONFIRM]
.
20. Verify the
CLOSE AND LOCK DOOR
message displays.
21. Close and lock the door.
22. Verify the
PRIME TUBING
message displays. Press
[NO]
.
23. Verify the
DOOR LOCKED
message displays.
24. Press
[START]
.
25. Verify the
PRESS START/PAUSE BUTTON TO STOP
message displays, and the infuser
is delivering fluid.
26. When the loading dose is complete, wait for the
INFUSER STOPPED
alarm to sound
(approximately 35 seconds) and verify the
INFUSER STOPPED
message displays.
27. Open the door and verify the
STOPPED CONTINUOUS
message appears.
28. Select
CHANGE RX
.
29. Verify the
SELECT SETTING TO CHANGE
message displays.
30. Select
LOADING DOSE
, using the numeric keypad enter 10 mg then press
[ENTER]
.
31. Verify the
SELECT SETTING TO CHANGE
message displays.
32. Select
SAVE AND EXIT
.
33. Verify the
LOADING DOSE
message displays. Press
[CONFIRM]
.
34. Verify the
CONT RATE
message displays. Press
[CONFIRM]
.
35. Verify the
CLOSE AND LOCK DOOR
message displays.
36. Close and lock the door.
37. Verify the
DOOR LOCKED
message displays.
38. Press
[START]
.
39. Verify the
PRESS START/PAUSE BUTTON TO STOP
message displays, and the infuser
is delivering fluid.