5.7 NETWORK CONNECTIVITY CHECKS
Technical Service Manual
5 - 45
LifeCare PCA
4. If the
UNABLE TO AUTO-PROGRAM
message is displayed and at least two minutes
have passed since the infuser was powered on, then the infuser is not connected
to your ICU Medical MedNet server
.
Figure 5-21. Unable to Auto-Program Screen
If the infuser is not configured for Auto Programming, the status of the connection to the
MedNet server and detailed information on the network configuration may be reviewed in
Diagnostic mode
.
If connection problems occur, contact the local IT representative or
.
PRESS MANUAL
PROGRAM KEY
TO CONTINUE
MANUAL
PROGRAM
Morphine
5 mg/mL
CCA1
UNABLE TO
AUTO-PROGRAM