SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 6
Technical Service Manual
To clean the back of the infuser:
1. Turn the infuser upright to view the back
.
Figure 5-2. Rear View - Cleaning the Back of Infuser
2. Using lint-free cloths dampened in the approved cleaning solution or wipes, wipe
the infuser handle and back surface.
3. Using a combination of brushing, swabbing, and wiping (either with lint-free cloth
dampened in the approved cleaning solution or wipes), clean the threaded pole
clamp shaft by rotating the knob back and forth, and clean the knob highlighted
in
.
4. Use a brush to remove debris and then moist swabs to clean any openings or hard
to reach areas highlighted in
Note:
Do not allow fluid into any ports or vents on the infuser.
Note:
Do not allow fluid into flow behind the pole clamp knob.
5. Using a lint-free cloth dampened in the approved cleaning solution or wipes, wipe
areas around the power cord outlet, the power cord, and the plug, including prongs.
6. Using a lint-free cloth dampened in the approved cleaning solution or wipes, wipe
areas around the patient pendant, pendant cord, and the button at the end of the
pendant cord.
Note:
Do not allow fluid to flow into the crevices near the button.