SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 32
Technical Service Manual
Figure 7-12. Rear Enclosure External Components
7.3.9.1
FUSE
REPLACEMENT
Recommended tools for this procedure are a No. 2 Phillips screwdriver and a small
flat blade screwdriver.
The replacement part for this procedure is:
Fuse, 1.6 A, 250 V, Slo-Blo
To replace the fuses, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Using the Phillips screwdriver, remove the screw that attaches the battery door
to the rear of the infuser.
3. To access the fuses, remove the AC power cord from its receptacle by grasping
the plug. Do not pull on the cord.
4. Locate the plastic fuseholder directly above the AC power receptacle.
Insert the flat blade screwdriver between the locking tab at the center bottom
of the fuseholder and the infuser housing. Press the tab toward the center
of the fuseholder to release it.
VELCRO
STRAP (2)
RETAINER
PLATE
6-32 x 1/4
FLAT HEAD
SCREW (2)
POLE CLAMP
ASSEMBLY
8-32 x 3/4
FLAT HEAD
SCREW (4)
4-24 x 3/8
THREAD
CUTTING
SCREW
EQUIPOTENTIAL
POST
FLAT
WASHER
M12 HEX NUT
ETHERNET
CONNECTOR
ETHERNET
LED INDICATORS
REAR
ENCLOSURE
4-40 x 3/8
FLAT HEAD
SCREW (2)
FUSE (2)
BATTERY
STRAP
#6 FLAT
WASHER