SECTION 1 INTRODUCTION
LifeCare PCA
1 - 10
Technical Service Manual
8. Enter the Access Code and press
[CONTINUE]
. The current Alarm Volume displays,
flashing.
Note:
To obtain the Access Code,
9. Select the desired Alarm Volume, and confirm that the desired volume is flashing.
10. Press
[EXIT]
to save the flashing Alarm Volume and return to the
BIOMED MODE
screen.
1.10.6
CLEANING
PRIOR
TO
FIRST
USE
Before placing the PCA Infuser in service, follow the cleaning instructions in
through
.
1.11
SUSPECTED
CYBERSECURITY
EVENT
OR
THREAT
The section contains information on the recommended procedure upon detecting a
suspected cybersecurity event or threat.
1. Contact hospital and/or follow hospital guidelines to report the suspected
cybersecurity event or threat.
Attempts to exploit a remote vulnerability on an infusion device would require
penetration of several layers of network security enforced by the hospital, including
firewalls. These measures serve as the primary defense against tampering with a
medical device.
2.