SECTION 6 TROUBLESHOOTING
LifeCare PCA
6 - 8
Technical Service Manual
489
User Interface Data Error
Failed CRC of saved user
interface data
For each malfunction
press
[CLEAR]
,
then
[ON/OFF]
to power
off the infuser,
then
[ON/OFF]
to power
on the infuser;
continue until all
malfunctions are cleared
Verify lithium coin cell
battery connections
Verify infuser is connected
to AC power
Replace lithium coin cell
battery
Replace battery
If malfunction cannot be
cleared,
490
Error Log Data Error
Failed error log checksum
491
Background Data Error
Inconsistent background
data
492
Switch Data Error
Corrupted switch data
493
Mirror Data Mismatched
Invalid date and time
494
UI State Error
User interface state ID
out of range
495
Backlight Data Error
Invalid display data
Replace LCD assembly
If problem persists,
496
Vial Manager Data Error
Failed CRC of vial manager
data
497
Power Data Error
Failed CRC of battery
charging data
498
Display Data Error
Invalid display data
Replace LCD assembly
If problem persists,
Table 6-2. Error Codes Requiring Technical Service
Alarm
Code
Malfunction Description
Possible Cause
Corrective Action