SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 12
Technical Service Manual
7.3.8.1
SECURITY
DOOR
AND
DOOR
HINGE
REPLACEMENT
The recommended tool for this procedure is a No. 1 Phillips screwdriver.
Replacement parts for this procedure are:
Door, Security
Hinge, Security Door
Screw, 4-40 x 3/8, Pan Head, Phillips
To replace the security door and door hinge, see
,
and proceed as follows:
1. Disconnect the infuser from AC power.
2. Unlock the security door, and open the door approximately 270 degrees to reveal
the four screws that attach the hinge.
3. Using the Phillips screwdriver, remove the four 4-40 screws and remove the hinge
and the door.
4. Attach the replacement door and/or hinge using the screws that were removed
in Step 3.
5. Close and lock the security door. Verify that the door aligns with the door latch
assembly and that the door is locked.
Replacement of the security door and door hinge is routine maintenance and no verification
procedure is normally required. However, if the infuser may have been damaged during
the procedure, perform the PVT in