SECTION 5 MAINTENANCE AND SERVICE TESTS
LifeCare PCA
5 - 36
Technical Service Manual
Figure 5-16. Vial/Injector Position for Stuck Injector Switch Pin Test
7. If no stuck switch injector pin condition exists, the
CHECK INJECTOR
message
displays. The test is complete; press
[ON/OFF]
to power off the infuser and continue
the PVT.
8. If the infuser attempts to scan the vial's barcode with the injector flange disengaged
from the injector flange retainer, then a stuck injector switch pin condition exists.
Perform the procedure for cleaning the injector switch pin in
, then
repeat this test.
Note:
If repeating the Stuck Injector Switch Pin Test results in failure after the
cleaning procedure in
has been performed, replace the
mechanism assembly as described in
or
.
5.6.4
SELF
TEST
Conduct all tests with the infuser connected to AC power unless otherwise specified.
Note:
If the quality of earth grounding source is in doubt, use battery power.
If an alarm condition occurs during the self test, press
[ON/OFF]
twice. If the alarm
condition recurs, note the message and take corrective action
.
Repeat the self test. If the alarm condition continues to recur, remove the infuser from
service and
CAUTION:
Do not place the infuser in service if the self test fails.