
APPENDIX
LifeCare PCA
A - 4
Technical Service Manual
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the infuser
is used exceeds the applicable RF compliance level above, the infuser should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the infuser.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
1
] V/m.