Technical Service Manual
7 - 1
LifeCare PCA
Section
7
REPLACEABLE
PARTS
AND REPAIRS
This section contains adjustment procedures, safety and equipment precautions,
required tools and materials, and step-by-step procedures for replacing parts
in the infuser.
Unless otherwise stated, always perform the PVT after a replacement procedure.
Figures are rendered as graphic representations to approximate actual product. Therefore,
figures may not exactly reflect the product.
7.1
REPLACEABLE
PARTS
Replaceable
parts for the infusion system are itemized in the Illustrated Parts Breakdown
(IPB) and are identified in
and
.
identifies each part by an
index number that correlates to
To view the online replacement parts list, visit the website at
www.icumed.com
.
7.2
ADJUSTMENT
PROCEDURES
Before opening the front enclosure of the infuser, take all necessary precautions
for working on high-voltage equipment.
Unless otherwise indicated, the following procedures require an empty syringe
to be installed in the cradle.
WARNING: EXPLOSION HAZARD EXISTS IF THE INFUSER IS SERVICED
IN THE PRESENCE OF FLAMMABLE SUBSTANCES.
WARNING: UNLESS OTHERWISE INDICATED, DISCONNECT THE INFUSER FROM AC
POWER BEFORE PERFORMING ADJUSTMENT OR REPLACEMENT PROCEDURES.
CAUTION:
Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation.
Store PWAs in antistatic bags before placing them on any surface.