5-4
5
SRQ
ERR
SRQ
+
ERR
ERR
+
LSTN
ERR
+
LSTN
+
TALK
Solid
LED
PWR
(After
blinking
address)
RDY off
(After
selftest)
GPIB
Address
Setting
GPIB
CPU type
CPU type
Error
Program
hung
DCD low
Defective GPIB controller chip. Replace
GPIB Controller, U6 in 4899A, U3 in 4809A
or U8 in 4819A.
Address value should be between 0 and 30.
Check and or correct address setting.
Wrong firmware or GPIB chip for hardware
configuration. Check hardware configuration
and change firmware or replace GPIB Control-
ler chip with NI 7210. U8 in 4819A
Wrong CPU type. Should be Zilog
Z8S18020VSC (SL19A19A Enhanced Ver-
sion)
Wrong CPU type. Should be Zilog
Z8S18020VSC
Possible Fault
Open GPIB chip selection line or grounded inter-
rupt into Z180.
DCD input should be high. Check cable wiring
and slave device is handshaking signals used in
cable.
TABLE 5-1 SELF TEST ERROR CODES
CONTINUED
Blinking
LED
Error
Possible Fault