7-4
ELISYS QUATTRO
User Manual HUMAN
Error Messages
Reading at %1/%2
Message: measurement at wave-
lengths "%1"/"%2".
-
Reagent ... is unde-
fined
The reagent has not been defined.
Reagent has to be defined in the rea-
gent database.
Enter the parameters of the reagent in
the reagent database.
Resuming after a
crash
PC does not respond while process-
ing the worklist and has to be
restarted.
Following restart of the PC, you may be
able to save the worklist and the results.
Connection to the analyzer is estab-
lished and the results stored there are
transferred to the PC.
RS232 write error -
command: '%1'
Error in interface PC / analyzer.
Check connections.
Shaking at %1 for %2
seconds
Message indicating that shaking is
performed at a frequency of %1 for
the duration of %2 seconds.
-
Some required
resources have not
been allocated to sys-
tem positions
Not all required reagents have been
allocated to a position.
All reagents requested in the Load
window have to be allocated to a posi-
tion or have to be loaded.
Suspect tip pickup
Pipettor cannot take up pipettor tip
correctly (does not move deep
enough into the tip).
New teaching of pipettor required. Call
service.
System cover closed
Message instrument cover is closed. -
System cover opened
Message instrument cover is
opened.
Close cover, otherweise the system can
not start processing a worklist.
System fluid low
System fluid for pipettor too low.
Fill up system fluid container with de-
ionized water.
The plate transport
cannot find a route
from its current posi-
tion to its next desti-
nation
Plate transport has no information
about its present position.
Select the menu item Utilities |
Selftest to initialise the system
again. If the error persists, call service.
Tip eject failure
Error in the tip eject station.
Remove tip manually from the pipettor
or trigger the eject mechanism manu-
ally. If this error recurs, call service.
Unknown colorimeter
error code %1
Unknown photometer error.
Call service.
Unknown incubator
error code %1
Unknown incubator error.
Call service.
Unknown plate trans-
port error code %1
Unknown plate transport error.
Call service.
Unknown washer
error code %1
Unknown washer error.
Call service.
Error:
Cause:
Action:
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...