Technical Data
ELISYS QUATTRO
User Manual HUMAN
8-1
8
Technical Data
System Overview
Number of plates
Up to 4 plates at once
Number of samples
Max. 200 tubes,
16 mm
Number of reagents Depending on rack configuration
Controls & Standards
Depending on rack configuration
Dilution area
For up to 152 dilution tubes
Number of tips
Up to 5 tip racks for 300
l or 1100
l disposable
tips
Loading
Continuous loading of plates, samples, reagents
and tips
Photometer
Spectral range
400 - 700 nm
Dynamic range
-0.100 to 3.000 O.D.
Accuracy
+/- 0.005 or 2.5 %
Linearity
0 - 2.000 O.D. +/- 1 %
Detection
Photo diode
Reading time
< 15 seconds
Read modes
OD
Filters
Up to 8 positions