Performing a Test Run
ELISYS QUATTRO
User Manual HUMAN
4-19
Action on error “Log and continue”
If a clot is detected, the system does not show a
message and continues with:
• The pipettor moves to the eject position to
remove the tip.
• Entry in the Eventlog.
• The next sample will be pipetted.
• The result file flags all samples with “Clot
detected” where a clot was detected.
Action on error “Manual pipette at end of step”:
If a clot is detected, the system does not show a
message and continues with:
• The pipettor moves to the eject position to
remove the tip.
• Entry in the Event log.
• The next sample will be pipetted.
• At the end of the pipetting step, a message
comes up and shows you all samples which were
not pipetted or were a clot was detected.
• The result file flags all samples with “MP” where a
clot was detected.