Error Messages
ELISYS QUATTRO
User Manual HUMAN
7-1
7
Error Messages
7.1
General Error Messages
If the
ELISYS QUATTRO
does not work, this is often due to minor problems which
you can deal with by yourself.
This chapter describes error messages and gives instruction on error recovery.
'%1' and '%2' are placeholders for a system module or the designation of a plate, a
reagent, or an error number.
System messages appear in the status bar of the
Phoenix
software, error messages
are displayed in a separate window, which has to be confirmed.
Error:
Cause:
Action:
ABORT button
pressed
Message: The Stop button has
been pushed during a run.
The run has been interrupted and may
be continued or aborted completely.
Aborting plate %1
The run of the plate has been
aborted. Reasons: The Stop but-
ton has been pushed.
The run has been interrupted and may
be continued or aborted completely.
Bar code IC error
The bar code cannot be read.
Verify the readability of the bar codes.
Check the bar code parameters in the
Setup menu. Try reading the bar-
codes once more. If these attempts fail,
call service.
Clot detected in rea-
gent %1
Clots were detected in reagent %1.
Abort the run, if necessary, and replace
reagent.
Colorimeter %1 not
homed
The X-motor of the photometer is not
in the home position.
Select the menu item Utilities |
Selftest to initialise the system
again. If the problem recurs, call serv-
ice. If this error message comes up dur-
ing a run, this run has to be aborted.
Colorimeter A/D error
Error of the analog/digital converter
of the photometer.
Call service.
Colorimeter A/D over
range error
Upper limit of photometer A/D has
been exceeded due to the signal
height of the preselected resolution.
The filter has not been installed prop-
erly. Call service.
Colorimeter A/D
under range error
Signal detection too low. Defect of
power supply unit of the photometer,
or something similar.
The halogen lamp of the colorimeter is
faulty and has to be replaced. If the
error persists, the optical components in
the photometer (filter, upper or lower
optic block) may be dirty. Call service.
Colorimeter back-
ground light level
error
Typically occurs when light entered
the measurement chamber.
Avoid direct exposure to sunlight, close
all housing doors and covers.
Select the menu item Utilities |
Selftest. to initialise the system
again.
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...