8-2
ELISYS QUATTRO
User Manual HUMAN
Technical Data
Pipetting System
Pipettor
Liquid pipettor for disposable tips
Liquid level detec-
tion
Standard (capacitive)
Min.-Max. volumes
10
l to 300
l with 300
l tips or
301
l to 1000
l with 1100
l tips
Accuracy
< -15 % at 25
l (aqua)
< -5 % at 100
l (aqua)
Precision
< 5 % CV at 25
l
< 2.5 % CV at 100
l
Features
Tip detection, mixing, multi-dispensing
Washing
Capacity
Up to 4 buffers
Wash head
1 x 8
Dispense volume
200 - 2500
l/well
Precision
+/- 5 % CV at 300
l
Residual volume
< 2.5
l in U-shaped bottom
< 4
l in flat bottom
Fluid alarms
Low reagent, waste full
Features
Sweeps, soak, purge, top and bottom wash, vari-
able pump speeds
Others
Bar code
Samples and in rack system
Plate bar code read-
er
Optional
Interfaces
ASTM interface
Time Scheduling
Schedules 4 plates
Multiple assays per
plate
Up to 12 assays per plate
Export options
ASCII export and report
Qualitative
User definable result classes
Quantitative
Linear to quadratic regression, sigmoidal, and
many more
Language support
Optional; supported via use of Passolo (commer-
cially available program)
QA analysis
Mean, SD, CV, Standard Error and Levey-Jen-
nings (basic version)
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...