7-2
ELISYS QUATTRO
User Manual HUMAN
Error Messages
Colorimeter EEPROM
error
Error during EEPROM reading or
writing.
Call service.
Colorimeter filter %1
error
Error at filter no. %1. The gain factor
for the respective filter cannot be
identified. Filter is faulty or wrong fil-
ter has been installed.
Call service to check the installed filter
and to replace it, if necessary.
Colorimeter filter
motor home error
The system does not recognise the
current position of the filter motor.
Move to the home position.
Select the menu item Utilities |
Selftest to initialise the system
again.
Colorimeter filter
motor movement
error
The system does not recognise the
current position of the filter motor.
Move to the home position.
Select the menu item Utilities |
Selftest to initialise the system
again.
Colorimeter lamp
error
Halogen lamp of photometer is
faulty.
Replace the halogen lamp.
Colorimeter optic
channel %1 error
Error in one of the 9 optical chan-
nels. Upper or lower optic block is
dirty or faulty.
The optical blocks have to be cleaned
by service personnel. Call service.
Colorimeter plate
motor home error
The system does not recognise the
current position of the plate motor in
the measuring area. The plate motor
has to be moved to the home posi-
tion.
Select the menu item Utilities |
Selftest to initialise the system
again.
Colorimeter plate
motor movement
error
Error of motor when moving the
plate to the measurement area.
Select the menu item Utilities |
Selftest to initialise the system
again. If the error persists, call service.
Colorimeter voltage
reference error
Error in reference voltage of the pho-
tometer.
Call service.
COMGEN error '%1'
RS232 connection faulty.
Check correct connection between PC
and analyzer. Start PC and analyzer
again.
COMGEN target over-
flow error
-
Call service.
COP serial port test
error
The components are connected via
serial interfaces to the control board.
Error in serial interfaces.
Call service.
ERROR: Argument
error in '%1'
Component cannot be actuated.
Error in component %1.
Call service.
Error scheduling
plate '...'
Error in assay or missing resources.
This error message is displayed in
the worklist for the respective plate,
after the active worklist has been
verified.
The worklist or the assay has to be
changed accordingly.
Incubator heater %1
error
Defect in heater foil no. %1.
Call service.
Incubator sensor %1
error
Defect of incubator sensor no. %1.
Call service.
Error:
Cause:
Action:
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...