4-16
ELISYS QUATTRO
User Manual HUMAN
Performing a Test Run
4.4.4
Reagent Volume Check
Do not change any reagent positions!
After loading and allocation of all required resources,
press on the OK button. A volume check of all rea-
gents is carried out automatically.
If the reagent check shows that the volume of one
reagent is low, a System Error dialogue box
appears alerting you about any insufficient volume.
To refill the respective reagent or to replace it by a
new bottle, the respective reagent rack has to be
moved out, reagent has to be refilled or the bottle
replaced, respectively and the rack has to be
inserted again.
• If you press on the Refill Bottle... button,
the volume is checked once more.
If the reagent was identified manually, the posi-
tion in the rack has to be allocated once more
(see above).
• If the reagent cannot be supplied, press on the
Abort Worklist button.
This will abort any further volume check.
• The Ignore button skips the reagent check for
this specific reagent (the insufficient volumne is
logged) and continues with the reagent check for
the next one.