4-24
ELISYS QUATTRO
User Manual HUMAN
Performing a Test Run
Allocate the vials only into suitable locations. An incorrect allocation may
cause errors in dispensing!
When refilling a reagent bottle never exceed the bottle shoulder. An over-fill-
ing may lead to an incorrect dispensing!
Never remove a rack while still in use. Remove it only when the related LED
is flashing. If a sample rack is removed which is still in use, all samples not
already pipetted out of this rack will be flagged with
Removed
and not cal-
culated.
Make sure that the flap of the reagent unit is closed after re-loading.
6.
Patient samples
:
Insert sample racks and allocate sample symbols
with the mouse to the new sample rack
Processed sample racks (flashing red LED) can
be removed.
7.
Reagents:
Refill or add additionally required reagents as
shown on the screen
(see chapter 4.4.2 on page 4-12)
.
The reagent symbols are automatically allocated
to the respective positions.
8.
Disposable tips and dilution plates
:
Refill disposable tips, dilution plates, wash buffer
and distilled water as shown on the screen
9. After loading and allocation of all required
resources confirm the dialogue with OK. A vol-
ume check of all reagents is carried out automati-
cally
(see chapter 4.4.4 on page 4-16)
.
After the volume check the worklist is calculated new
(interlacing or adding new plates/assays to be proc-
essed). Further processing is then carried out in
accordance with the new worklist.
Summary of Contents for Elisys Quattro
Page 1: ...HUMAN ELISYS QUATTRO User Manual Elisys Quattro User Manual Cat No 16300 1...
Page 6: ...TOC 4 ELISYS QUATTRO User Manual HUMAN...
Page 26: ...2 12 ELISYS QUATTRO User Manual HUMAN System Basics...
Page 66: ...5 8 ELISYS QUATTRO User Manual HUMAN Loading Unloading Disposal...
Page 69: ...Service and Maintenance ELISYS QUATTRO User Manual HUMAN 6 3 Pipettor...
Page 87: ...Technical Data ELISYS QUATTRO User Manual HUMAN 8 5...