203
Equivalent Front-Panel Key Commands
To Set the Display Units for High and Low Measurement Limits.
Use the
:HLIMit:DUNits and :LLIMit:DUNits commands to set the units-of-measure used
to display the high and low measurement limit values. Refer to
for description of Display Units.
Syntax
:HLIMit:DUNits <disp_units>
:LLIMit:DUNits <disp_units>
Example
OUTPUT 714;"MEAS:AFR:FM:HLIM:DUN KHZ"
OUTPUT 714;"MEAS:AFR:FM:LLIM:DUN KHZ"
This sets the high and low measurement limit Display Units to kHz for
the Audio Frequency Analyzer FM Deviation measurement.
NOTE:
When querying measurement limits through HP-IB, the Test Set always returns
numeric values in Attribute Units, regardless of the current Display Units or
HP-IB Units settings. Numeric values are expressed in scientific notation. Refer
to
"Attribute Units (AUNits)" on page 73
To Query the Display Units for High and Low Measurement Limits.
Use the
:HLIMit:DUNits? and :LLIMit:DUNits? commands to query the units-of-measure
used to display the high and low measurement limit values. Refer to
for description of Display Units.
Syntax
:HLIMit:DUNits?
:LLIMit:DUNits?
Example
OUTPUT 714;"MEAS:AFR:FM:HLIM:DUN?"
ENTER 714;Hi_disp_unit$
OUTPUT 714;"MEAS:AFR:FM:LLIM:DUN?"
ENTER 714;Lo_disp_unit$
This queries the high measurement limit Display Units for the Audio
Frequency Analyzer FM Deviation measurement.
NOTE:
When querying measurement limits through HP-IB, the Test Set always returns
numeric values in Attribute Units, regardless of the current Display Units or
HP-IB Units settings. Numeric values are expressed in scientific notation. Refer
to
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...