657
Error
−
110
Command header error
An error was detected in the header.
Error
−
111
Header separator error
A character which is not a legal header separator was encountered while parsing the
header.
Error
−
112
Program mnemonic too long
The header contains more than twelve characters (see IEEE 488.2,7.6.1.4).
Error
−
113
Undefined header
The header is syntactically correct, but it is undefined for this specific device. For
example, *XYZ is not defined for any device.
Error
−
114
Header suffix out of range
Indicates that a nonheader character has been encountered in what the parser expects is a
header element.
Error
−
120
Numeric data error
This error, as well as errors
−
121 through
−
128, are generated when parsing a data element
which appears to be numeric, including the nondecimal numeric types.
Error
−
121
Invalid character in number
An invalid character for the data type being parsed was encountered. For example, an
alpha in a decimal numeric or a “9” in octal data.
Error
−
123
Exponent too large
The magnitude of the exponent was larger than 32000 (see IEEE 488.2, 7.7.2.4.1).
Error
−
124
Too many digits
The mantissa of a decimal numeric data element contained more than 255 digits excluding
leading zeros (see IEEE 488.2, 7.7.2.4.1).
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...