Chapter 1, Using HP-IB
Addressing
Addressing
Factory Set Address
The Test Set’s HP-IB address is set to decimal 14 at the factory. The address can be
changed by following the instructions in
"Setting the Test Set’s Bus Address" on page
.
Extended Addressing
Extended addressing (secondary command) capability is not implemented in the Test
Set.
Multiple Addressing
Multiple addressing capability is not implemented in the Test Set.
Setting the Test Set’s Bus Address
The Test Set’s HP-IB bus address is set using the
HP-IB Adrs
field which is located
on the I/O CONFIGURE screen. To set the HP-IB bus address; select the I/O
CONFIGURE screen and position the cursor next to the
HP-IB Adrs
field. The
address can be set from decimal 0 to 30 using the numeric DATA keys, or by
pushing and then rotating the Cursor Control knob. There are no DIP switches for
setting the HP-IB bus address in the Test Set. The new setting is retained when the
Test Set is turned off.
Displaying the Bus Address
The Test Set’s HP-IB bus address can be displayed by pressing and releasing the
SHIFT key, then the LOCAL key. The address is displayed in the upper left-hand
corner of the display screen.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...