Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Reading-back Attribute Units.
Use the AUNits? command to read back the Attribute Units setting for the
selected measurement. For example, the following BASIC language program
statements show how the AUNits? command can be used to read-back a
Distortion REFerence SET level:
!Query the REFerence SET value for the Distortion measurement
OUTPUT 714;"MEAS:AFR:DIST:REF:VAL?"
!Read the REFerence SET value into variable Ref_set_val
ENTER 714;Ref_set_val
!Query the Attribute Units setting for the Distortion measurement
OUTPUT 714;"MEAS:AFR:DIST:AUN?"
!Read the Attribute Units setting into string variable Atribute_set$
ENTER 714;Atribute_set$
!Print out the variables in the form <VALUE><UNITS>
PRINT Ref_set_val;Atribute_set$
If a reference of 25% is set, 25 PCT would be printed.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...