28
Chapter 1, Using HP-IB
Getting Started
General HP-IB Programming Guidelines
The following guidelines should be considered when developing programs which
control the Test Set through HP-IB:
•
Guideline #1. Avoid using the TX TEST and RX TEST screens.
The RX TEST and TX TEST screens are specifically designed for manual testing of
land mobile FM radios and, when displayed, automatically configure six “priority”
fields in the Test Set for this purpose. The priority fields and their preset values are
listed in
. When the TX TEST screen or the RX TEST screen is dis-
played, certain priority fields are hidden and are not settable. The priority fields which
are hidden are listed in
When the TX TEST screen or the RX TEST screen is displayed, any HP-IB commands
sent to the Test Set to change the value of a hidden priority field are ignored. Hidden pri-
ority fields on the TX TEST or RX TEST screens are not settable manually or program-
matically.
Displaying either of these screens automatically re-configures the 6 “priority” fields as
follows:
1. When entering the RX TEST screen,
a. the RF Generator’s
Amplitude
field, the
AFGen1 To
field and the AF An-
alyzer’s measurement field (measurement displayed in upper, right portion of
CRT display) are
•
set to their preset values upon entering the screen for the first time since
power-up, OR
•
set to their preset values if the PRESET key is selected, OR
•
set to the last setting made while in the screen
b. the RF Generator
Amplitude
field and the
AFGen1 To
field are
•
set to their preset values whenever entering the screen, OR
•
set to their preset values if the PRESET key is selected
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...