376
Chapter 6, Memory Cards/Mass Storage
Using External Disk Drives
Using External Disk Drives
The Test Set supports only HP-IB external disk drives. Certain configuration
information is required by the Test Set to access external disk drives.
The I/O CONFIGURE screen’s HP-IB
Mode
field must be set to Control any time
an external disk drive is used by the Test Set.
To load files from the TESTS screens or NMT Signaling Decoder screen, the
disk’s mass storage volume specifier must be entered in the
External Disk
Specification
field on the TESTS (External Devices) screen (for example,
:,702,1
).
Initializing External Disks
All new external disk media must be initialized before it can be used to store
information. External disk media can be initialized for either LIF (Logical
Interchange Format) or DOS (Disk Operating System) format using the Test Set.
(See
"DOS and LIF File System Considerations" on page 358
.)
External disk media can be initialized from the TESTS (IBASIC Controller)
screen by inserting the new media into the external disk drive and executing the
following IBASIC command:
INITIALIZE "<volume type>:<external disk mass storage volume
specifier>"
where the <volume type> can be LIF or DOS
For example:
INITIALIZE "DOS:,702,1").
To verify that disk media has been properly initialized, execute the IBASIC
command:
CAT "<external disk mass storage volume specifier>"
For example:
CAT ":,702,1"
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...