Chapter 8, Programming The Call Processing Subsystem
CALL BIT Screen Message Field Descriptions
CPA
Combined paging/access field.
1 binary character required.
CMAX-1
CMAX is the number of access channels in the system.
7 binary characters required.
END
End indication field.
•
Set to 1 to indicate the last word of the overhead message train.
•
Set to 0 if not the last word.
1 binary characters required.
OHD
This field displays the overhead message type.
•
A ‘100’ indicates a global action message.
•
A ‘110’ indicates that this is the first word of the system overhead parameter message.
•
A ‘111’ indicates this is the second word of the system parameter overhead message.
3 binary characters required.
Parity
Parity field. The contents of the Parity field cannot be set by the user. The Test Set calculates the
parity bits.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...