513
Chapter 8, Programming The Call Processing Subsystem
Programming The AUTHENTICATION Screen
ESN
This field is used to enter the electronic serial number (ESN) of the mobile station
into the Test Set. There are two methods which can be used to enter the
ESN
. The
serial number can be captured by the Test Set through a registration order, or the
serial number can be entered as an 8-digit hexadecimal value directly into this
field.
To capture the electronic serial number through a registration , see
for further information.
To enter the serial number directly use the :ESNumber command.
The query form of the command (that is, :ESNumber?) can be used to determine
the current setting of the ESN field.
Syntax
:ESNumber <’><8 character hex value><’>
:ESNumber?
Example
OUTPUT 714;"CALLP:AMPS:AUTH:ESN ’12D4E678’"
OUTPUT 714:"CALLP:AMPS:AUTH:ESN?"
ENTER 714;Esn_val$
RAND_A
RAND_A is the 16 most significant bits of
RAND
and is entered as a 4-digit
hexadecimal value. The mobile station stores and uses the most recent version of
RAND_A in the authentication process. RAND_A is user definable.
The :RAND:A command is used to control this field.
The query form (that is, :RAND:A?) can be used to query the current field value.
Syntax
:RANDA
:RANDA?
Example
OUTPUT 714;"CALLP:AMPS:AUTH:RAND:A ‘4F5A’"
OUTPUT 714;"CALLP:AMPS:AUTH:RAND:A?"
ENTER 714;Rand_val$
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...