Chapter 8, Programming The Call Processing Subsystem
Programming The ANALOG MEAS Screen
Example Measurement Routines
There are a wide variety of audio measurements which can be made from the
ANALOG MEAS screen.
The following examples illustrate how to make a typical mobile station receiver
measurement (RF Sensitivity) and a typical mobile station transmitter
measurement (FM Hum and Noise).
Refer to the HP 8920A RF Communications Test Set Applications Handbook,
section “Testing FM Radios” for further information on using the Test Set’s
Audio Analyzer to make audio measurements.
Example RF Sensitivity Measurement
The following example code segment shows how to program the
ANALOG MEAS
screen to make an RF Sensitivity measurement. The code segment represents a
HP RM BASIC subprogram.
In order for this subprogram to work properly the following conditions must be
true when the subroutine is called:
•
Call Processing Subsystem is in the
Connect
state (that is, the
Connect
annunciator
is lit)
•
the mobile station’s speaker output is connected to the Test Set’s AUDIO IN connector
•
the mobile station’s microphone input must be connected to the Test Set’s AUDIO
OUT connector
The intended purpose of this example subprogram is to illustrate how to program
the
ANALOG MEAS
screen. There are a variety of ways to make an RF Sensitivity
measurement. The method used in this example is based upon the EIA/IS-19-B
Standard (May 1988). The method and standard chosen for any particular
application will depend upon the mobile station being tested.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...