433
Chapter 7, IBASIC Controller
PROGram Subsystem
:EXECute <program_command>
The :EXECute command is used to execute, from an
external controller, an IBASIC program command in the Test Set’s built-in IBASIC
Controller.
<program_command> is string data representing any legal IBASIC command. If the
string data does not represent a legal IBASIC command, an
IBASIC Error: -285
Program syntax error
is generated.
Any IBASIC program in the Test Set must be in either the PAUSed or STOPped state
before the external controller issues the :EXECute <program_command> command. If
the IBASIC program is in the RUN state, an
IBASIC Error: -284 Program
currently running
is generated.
Syntax
PROGram[:SELected]:EXECute <delimiter><program_command><delimiter>
The following notation is used in the command description:
<delimiter> = IEEE 488.2 <string data> delimiter, single quote or double quote, must be the
same.
Example
OUTPUT 714;"PROGram:SELected:EXECute ’CLEAR SCREEN’"
or
OUTPUT 714;"PROG:EXEC ’CLEAR SCREEN’"
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...