73
Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Attribute Units (AUNits)
Attribute Units are the units-of-measure used by the Test Set when sending or
receiving numeric data through HP-IB for the MEASure commands: REFerence,
METer (HEND, LEND, INT), HLIMit and LLIMit (refer to
Measurement Syntax" on page 193
for further details). These measurement
commands are analogous to the front-panel Data Function keys: REF SET,
METER, HI LIMIT and LO LIMIT respectively. Attribute Units use the same set
of units-of-measure as the HP-IB Units (except Frequency Error), but are only
used with the MEASure commands: REFerence, METer (HEND, LEND, INT),
HLIMit and LLIMit.
lists the Attribute Units used in the Test Set.
Default Data Function Values.
The majority of measurements made with the Test
Set can be made using the Data Functions: REF SET, METER, AVG, HI LIMIT
and LO LIMIT. Measurements which can be made using the Data Functions have
a black bubble with the comment “See Number Measurement Syntax” in their
syntax path. If one or more of the Data Functions are not available to that
measurement, the Data Function(s) not available will be listed under the black
bubble (see the
Measure syntax diagram, on page 163
Table 9
Attribute Units
Parameter
Unit of Measure
Power
Watts (W) or dBm (DBM)
Amplitude
Volts (V)
Frequency
Hertz (Hz)
Time
Seconds (S)
Data Rate
Bits per second (BPS)
Current
Amperes (A)
Resistance
Ohms (OHM)
Relative Level
decibels (DB) or percent (PCT)
Marker Position
Division (DIV)
FM Modulation
Hertz (HZ)
AM Modulation
Percent (PCT)
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...