330
Chapter 5, Advanced Operations
Instrument Initialization
The *RST Reset condition in the Test Set was specifically designed to configure
the instruments for manual testing of an FM radio. The *RST Reset default
display screen is the RX TEST screen. Other operational characteristics are also
affected by the *RST reset as follows:
•
All pending operations are aborted.
•
Measurement triggering is set to TRIG:MODE:SETT FULL;RETR REP.
•
Any previously received Operation Complete command (*OPC) is cleared.
•
Any previously received Operation Complete query command (*OPC?) is cleared.
•
The Test Set’s display screen is in the UNLOCKED state.
•
The Power-up self-test diagnostics are not performed.
•
The Contents of the SAVE/RECALL registers are not affected.
•
Calibration data is not affected.
•
The HP-IB interface is not reset (any pending Service Request is not cleared).
•
All Enable registers are unaffected: Service Request, Standard Event, Communicate,
Hardware #1, Hardware #2, Operation, Calibration,Call Processing and Questionable
Data/Signal.
•
All Negative Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operation, Calibration,Call Processing and Questionable Data/Signal.
•
All Positive Transition Filter registers are unaffected: Communicate, Hardware #1,
Hardware #2, Operation, Calibration,Call Processing and Questionable Data/Signal.
•
The contents of the RAM memory are unaffected.
•
The contents of the Output Queue are unaffected.
•
The contents of the Error Queue are unaffected.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...