Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Specifying Units-of-Measure for Settings and Measurement Results
Numeric settings and measurement results in the Test Set can be displayed using
one or more units-of-measure (V, mV, mV, Hz, kHz, MHz…). When operating
the Test Set manually, the units-of-measure can be easily changed to display
measurement results and field settings in the most convenient format. HP-IB
operation is similar to manual operation in that the units-of-measure used to
display numeric data can be programmatically changed to the most convenient
form.
NOTE:
When querying measurements or settings through HP-IB, the Test Set always returns numeric
values in HP-IB Units or Attribute Units, regardless of the current Display Units setting. Refer
to
"HP-IB Units (UNITs)" on page 70
and
"Attribute Units (AUNits)" on page 73
further information.
There are three sets of units-of-measure used in the Test Set: Display Units,
HP-IB Units, and Attribute Units. Writing correct HP-IB programs requires an
understanding of how the Test Set deals with these different sets of units-of-
measure.
Display Units (DUNits)
Display Units are the units-of-measure used by the Test Set to display numeric
data (field settings and measurement results) on the front-panel CRT display. For
example, the RF Generator’s frequency can be displayed in Hz, kHz, MHz and
GHz. Similarly, the measured TX Frequency can be displayed in Hz, kHz, MHz
and GHz.
When evaluating an entered value for a numeric field, the Test Set interprets the
data it receives in terms of the Display Units currently set. For example, if the
Display Units for the
RF Gen Freq
field are set to GHz and the operator tries to
enter 500 into the field, an
Input value out of range
error is generated
since the Test Set interpreted the value as 500 GHz which is outside the valid
frequency range of the Test Set.
Summary of Contents for 8920B
Page 10: ...10 ...
Page 16: ...Contents 16 9 Error Messages Index 673 ...
Page 17: ...17 1 Using HP IB ...
Page 48: ...48 Chapter 1 Using HP IB Remote Local Modes ...
Page 49: ...49 2 Methods For Reading Measurement Results ...
Page 61: ...61 3 HP IB Command Guidelines ...
Page 84: ...84 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Page 85: ...85 4 HP IB Commands ...
Page 90: ...90 Adjacent Channel Power ACP ...
Page 116: ...116 Configure I O Configure ...
Page 156: ...156 Call Processing ...
Page 185: ...185 System System SYSTem ERRor Returns integer value quoted string ...
Page 242: ...242 Common Command Descriptions ...
Page 251: ...251 5 Advanced Operations ...
Page 377: ...377 7 IBASIC Controller ...
Page 458: ...458 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Page 640: ...640 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Page 643: ...643 9 Error Messages ...
Page 672: ...672 ...
Page 682: ...Index 682 ...