Making Coordinated Measurements
When HP 64700 Card Cages are connected together via the Coordinated
Measurement Bus (CMB), you can start and stop up to 32 emulators at the
same time.
You can use the analyzer in one HP 64700 to arm (that is, activate) the
analyzers in other HP 64700 Card Cages or to cause emulator execution in
other HP 64700 Card Cages to break into the monitor.
You can use the HP 64700’s BNC connector (labeled TRIGGER IN/OUT on
the lower left corner of the HP 64700 rear panel) to trigger an external
instrument (for example, a logic analyzer or oscilloscope) when the analyzer
finds its trigger condition. Also, you can allow an external instrument to arm
the analyzer or break emulator execution into the monitor.
The coordinated measurement tasks you can perform are grouped into the
following sections:
•
Setting up for coordinated measurements.
•
Starting and stopping multiple emulators.
•
Driving trigger signals to the CMB or BNC.
•
Stopping program execution on trigger signals.
•
Arming analyzers on trigger signals.
304
Summary of Contents for 64782
Page 2: ...User s Guide for the Graphical User Interface HP 64782 MC6833x Emulation Analysis ...
Page 22: ...Part 1 Quick Start Guide 21 ...
Page 24: ...1 Getting Started 23 ...
Page 50: ...Part 2 User s Guide 49 ...
Page 52: ...2 Plugging into a Target System 51 ...
Page 54: ...3 Starting and Exiting HP 64700 Interfaces 53 ...
Page 64: ...4 Entering Commands 63 ...
Page 94: ...5 Configuring the Emulator 93 ...
Page 133: ...132 ...
Page 134: ...6 Using the Emulator 133 ...
Page 194: ...7 Using the Emulation Analyzer 193 ...
Page 255: ...254 ...
Page 256: ...8 Making Software Performance Measurements 255 ...
Page 290: ...9 Using the External State Analyzer 289 ...
Page 303: ...302 ...
Page 304: ...10 Making Coordinated Measurements 303 ...
Page 321: ...320 ...
Page 322: ...11 Setting X Resources 321 ...
Page 336: ...Part 3 Reference 335 ...
Page 337: ...Descriptions of the product in a dictionary or encyclopedia format Part 3 336 ...
Page 338: ...12 Emulator Analyzer Interface Commands 337 ...
Page 415: ...set Emulator Analyzer Interface Commands set 414 ...
Page 429: ... SYMB Emulator Analyzer Interface Commands SYMB 428 ...
Page 445: ...444 ...
Page 446: ...13 Error Messages 445 ...
Page 489: ...488 ...
Page 490: ...Part 4 Concept Guide 489 ...
Page 491: ...Topics that explain concepts and apply them to advanced tasks Part 4 490 ...
Page 492: ...14 Concepts 491 ...
Page 507: ...506 ...
Page 508: ...Part 5 Installation Guide 507 ...
Page 509: ...Instructions for installing and configuring the product Part 5 508 ...
Page 510: ...15 Installation 509 ...
Page 527: ...526 ...
Page 528: ...16 Installing Updating Emulator Firmware 527 ...
Page 535: ...534 ...
Page 539: ...538 ...
Page 549: ...548 ...