The
Run
Key
The
Run
key
causes
the
instrument
to
resume
acquiring
data.
If
the
instrument
is
stopped,
it
starts
acquiring
data
on
the
next
trigger
event.
If
the
instrument
is
already
in
the
Run
mode
,
it
continues
to
acquire
data
on
successive
trigger
events
.
If
pressing
the
Run
key
does
not
cause
waveform
data
to
display
on
the
screen,
try
the
following:
Press
the
A
utoscale
key
.
Make
sure
a
signal
is
connected
to
one
of
the
channels
and
the
display
for
that
channel
is
turned
on.
Make
sure
the
oset
does
not
have
the
trace
clipped
o
the
display
.
Check
the
trigger
setup
conditions
to
make
sure
the
trigger
conditions
are
valid
for
the
signal.
Set
the
Trigger
Sweep
mode
to
Freerun.
Freerun
forces
the
instrument
to
trigger
,
which
may
allow
you
to
see
enough
of
the
signal
so
that
you
can
set
up
the
front
panel
properly
.
2-7
Summary of Contents for 54750A
Page 1: ...User s Guide HP 83480A Analyzer HP 54750A Oscilloscope ...
Page 6: ...NOTE Clean the cabinet using a damp cloth only vi ...
Page 7: ...X Ray Radiation Notice vii ...
Page 8: ...Declaration of Conformity viii ...
Page 17: ...Figure0 1 Exampleof astatic safeworkstation xvii ...
Page 28: ......
Page 43: ...1 The Instrument at a Glance ...
Page 57: ...The Rear Panel Figure1 3 Theinstrument rearpanel 1 15 ...
Page 60: ...TheInstrument ataGlance ...
Page 61: ...2 General Purpose Keys ...
Page 69: ...3 Speci cations and Characteristics ...
Page 76: ...Speci cationsand Characteristics ...
Page 77: ...4 Calibration Overview ...
Page 84: ...CalibrationOverview ...
Page 85: ...5 Eye Mask and Eyeline Mode Measurement Tutorials ...
Page 136: ...Eye Mask andEyelineModeMeasurement Tutorials ...
Page 137: ...6 The Digital Communications Analysis Menus ...
Page 173: ...The Digital Communications Analysis Menus MeasureEyeMenu Figure6 7 Crossing measurement 6 37 ...
Page 178: ...The Digital Communications Analysis Menus MeasureEye Menu Figure6 11 MeasuringQ factor 6 42 ...
Page 191: ...7 Waveform Measurements ...
Page 211: ...8 Making Automatic Measurements ...
Page 233: ...9 Increasing Measurement Accuracy and Time Interval Measurement ...
Page 252: ...IncreasingMeasurement Accuracy andTime Interval Measurement ...
Page 253: ...10 General Purpose Oscilloscope Menus ...
Page 317: ...11 The General Function Menus ...
Page 345: ...The General Function Menus DisplayMenu Figure11 3 Connecteddots 11 29 ...
Page 415: ...12 Messages ...
Page 421: ...13 How the Instrument Works ...
Page 453: ...Index ...