General
Purpose
Oscilloscope
Menus
Histogram
Menu
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
Histogram
Scale
The
Histogram
Scale
softkey
gives
you
access
to
a
second-level
menu
that
allows
you
to
set
the
scale
of
the
histogram.
K
ey
Path
4
Histogram
5
NNNNNNNNNNNNNN
Mode
NNNNNNNNNNNNNNNNNNNNNNNNNN
waveform
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Hist
scale
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Scale
Type
Linear
sets
the
display
of
the
histogram
results
to
the
number
of
hits
per
division,
while
logarithmic
sets
the
display
of
the
histogram
results
to
dB
.
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
x
3
Linear
scale
F
or
linear
scale
type
,
the
scale
is
in
the
percentage
of
the
peak
per
division.
F
or
example
,
on
a
horizontal
histogram,
20%
places
one-fth
of
the
histogram
in
each
of
eight
divisions
with
the
top
of
the
peak
(100%)
at
the
middle
of
the
display
.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Log
scale
F
or
log
scale
type
,
the
scale
is
in
decibels
per
division.
The
histogram
is
plotted
according
to
the
following
formula.
dB
=
20
l og
X
peak
;
where:
X
is
the
number
of
hits
in
a
histogram
row
for
vertical
histograms
,
or
the
number
of
hits
in
a
histogram
column
for
horizontal
histograms
.
P
eak
is
the
number
of
hits
in
the
largest
histogram
column
or
row
.
dB
is
the
log
value
that
gets
plotted.
10-43
Summary of Contents for 54750A
Page 1: ...User s Guide HP 83480A Analyzer HP 54750A Oscilloscope ...
Page 6: ...NOTE Clean the cabinet using a damp cloth only vi ...
Page 7: ...X Ray Radiation Notice vii ...
Page 8: ...Declaration of Conformity viii ...
Page 17: ...Figure0 1 Exampleof astatic safeworkstation xvii ...
Page 28: ......
Page 43: ...1 The Instrument at a Glance ...
Page 57: ...The Rear Panel Figure1 3 Theinstrument rearpanel 1 15 ...
Page 60: ...TheInstrument ataGlance ...
Page 61: ...2 General Purpose Keys ...
Page 69: ...3 Speci cations and Characteristics ...
Page 76: ...Speci cationsand Characteristics ...
Page 77: ...4 Calibration Overview ...
Page 84: ...CalibrationOverview ...
Page 85: ...5 Eye Mask and Eyeline Mode Measurement Tutorials ...
Page 136: ...Eye Mask andEyelineModeMeasurement Tutorials ...
Page 137: ...6 The Digital Communications Analysis Menus ...
Page 173: ...The Digital Communications Analysis Menus MeasureEyeMenu Figure6 7 Crossing measurement 6 37 ...
Page 178: ...The Digital Communications Analysis Menus MeasureEye Menu Figure6 11 MeasuringQ factor 6 42 ...
Page 191: ...7 Waveform Measurements ...
Page 211: ...8 Making Automatic Measurements ...
Page 233: ...9 Increasing Measurement Accuracy and Time Interval Measurement ...
Page 252: ...IncreasingMeasurement Accuracy andTime Interval Measurement ...
Page 253: ...10 General Purpose Oscilloscope Menus ...
Page 317: ...11 The General Function Menus ...
Page 345: ...The General Function Menus DisplayMenu Figure11 3 Connecteddots 11 29 ...
Page 415: ...12 Messages ...
Page 421: ...13 How the Instrument Works ...
Page 453: ...Index ...